DOT Therapy: Tailored treatments, Improved resultsDOT2

DOT Therapy powered by the SmartXide DOT laser system employs a unique technology, SmartPulse, that allows simultaneous independent control of both energy (depth) and dwell time (thermal damage). This translates into a great benefit to the physician and the patient: Physicians can adjust parameters to tailor the treatment and downtime to the patient’s specific needs. Areas such as the neck, décolletage, eyelids, and darker skin types with a high risk of PIH (Postinflammatory Hyperpigmentation) – previously considered untreatable – can now be treated successfully and safely. Patients will enjoy the benefits of tailored treatments, reduced recovery time and quick procedures.

Control and Versatility

newdot-chart

With SmartXide DOT physicians can select treatment parameters ranging from superficial and delicate treatments (minimum density 1.7%) to increasingly deeper treatments such as skin resurfacing (density 100%). The HiScan scanner gives you a wide choice of patterns and sizes at your fingertips, and HiScan’s 38 density levels give you unparalleled precision.

Technological Innovation

Powerful technology in a streamlined system, the SmartXide DOT can deliver more ablative power to tissue, over 200 watts, than other lasers. The latest innovations in laser design and control systems lead to SmartPulse control of the treatment.  The adjustable high-power short-duration pulse to get to your desired ablation depth, coupled with the adjustable-duration low-power pulse to get the amount of thermal energy desired makes SmartXide DOT the most technically advanced CO2 system available.

SmartPulse technology makes SmartXide DOT the first system with independent
control of ablation depth and coagulation zone.

The HiScan™ Scanner

SmartXide DOT_HiScan Scanner                                                                 SmartXide DOT_HiScan Scanner patterns

The HiScan™ computer pattern generator puts controls at your finger tips. HiScan DOT meets all the traditional skin ablation needs by combining the benefits of both methods in one instrument, offering a selection of different scanning shapes, sizes, and aspect ratios.  It enables a higher degree of refinement in adapting the treatment to a patient.

Quality

A lightweight compact package, only 60 lbs, a precision touch controlled scanner, titanium-alloy articulated arm, a state-of-the-art pulsed laser source and a treatment driven touch screen control panel demonstrate quality, and experience of the world’s premier developer and manufacturer of medical CO2 lasers.

Customer Testimonials

 

Indications:

Incision, excision, ablation, vaporization and coagulation of body soft tissues including intraoral tissues, in medical specialties including aesthetic (dermatology and plastic surgery), otolaryngology (ENT), gynecology, neurosurgery, dental and oral surgery and genitourinary surgery. The use with the scanning unit is indicated for ablative skin resurfacing.

Specifications:

Source CO2
Wavelength 10.6 µm
Repetition Rate from 5 to 100 Hz
Thermal Control up to 2 ms
Pulse Length from 0.2 ms to 80 ms
Ablative Density 1.7% to 100%
Transmission system Titanium Alloy Articulated arm with 7 mirrors
Electrical Requirements 115 Vac/3.7 A /50-60 Hz
Dimensions 120 cm (H), 48 cm (L), 55 cm (D)
Weight 83 lb/38 kg (30W model)